DC Current Sensor
AC Current Sensor
Measuring Instruments
SPECIFICATIONS | |
---|---|
Withstanding Voltage | AC 3700V 1 minute max. (Between the core of CT and outer case) |
Storage Temperature | -10℃ ~ 60℃ < 70%RH( without condensation ) |
Accuracy | (23℃±5℃, 80%RH or less) Range/Resolution/Accuracy AC 40A/0.01A/±1.5%rdg±5dgt(50/60Hz) AC 400A/0.1A/±1.5%rdg±8dgt(50/60Hz) DC 40A/0.01A/±1.5%rdg±3dgt DC 400A/0.1A/±1.5%rdg±5dgt AC 60V/600V/0.01V/±1.0%rdg±8dgt DC 60V/600V/0.01V/±1.0%rdg±6dgt 1000Ω/0.1Ω/±1.5%rdg±8dgt |
Operating Temperature | 0℃ ~ 40℃ < 80%RH ( without condensation ) |
Power supply | 1.5V (AM-4, LR03 or AAA)×3 |
Power consumption | Approx. 9mA |
Weight | Approx. 135g |
Measuring Range | AC Current:40A/400A DC Current:40A/400A AC Voltage:60V/600V DC Voltage:60V/600V Resistance:1000Ω |
AC Conversion | True rms responding |
Safety standard | IEC61010-1, IEC61010-2-032 Installation CategoryⅡ 600V or CAT Ⅲ 300V |
Jaw opening capability | 30mmφ |
Limitation of circuit voltage | Less than AC/DC 600V |
Size | 45(W)×180(H)×24(D)mm |
Accessories | Soft case:1 Instruction manual:1 Test lead:1 Batteries:2 |
Crest factor | < 2.5 |
Display | 3.5 digit LCD max. reading of 3999 |
Over range indication | "OL" mark on LCD |
Data hold indication | "DH" mark on LCD |
Low battery indication | "B" mark on LCD |
Auto power off | Approx. 10 minutes later after power on |
E.M.C. standard | EN 61326 |
Measuring method | Dual slope integration mode |
Sampling time | 2 times/sec |
Battery life | Approx. 90 hours continuous |
Polarity | "-" sign for negative polarity |
DC current zero offset | By push button switch |