DC Current Sensor
AC Current Sensor
Measuring Instruments

| SPECIFICATIONS 1 | ||||
|---|---|---|---|---|
| Range | 3mA | 30mA | 300mA | 20A | 
| Resolution | 0.001mA | 0.01mA | 0.1mA | 0.01A | 
| Accuracy | ±1%rdg ±5dgt | |||
| SPECIFICATIONS 2 | |
|---|---|
| Measuring function | AC leakage current AC Load current  | 
| CT shape | Split clamp type circular ZCT | 
| CT inside diameter | 35mm | 
| Measuring frequency | 50 ~ 60Hz | 
| Sampling rate | 2 time/1s | 
| Measurement of temperature | Reference value displayed at “Multi-Tracer” | 
| Operation | Power ON/OFF (other function is at a paired unit) | 
| Low battery indication | Light up of LED on the main-body | 
| Applicable circuit voltage | AC600V | 
| Insulation resistance | More than 10MΩ by DC1kV insulation tester (Between CT~Universal connector)  | 
| Withstanding voltage | AC3.7kV/1min (Between CT and the Universal connector)  | 
| Operating Temperature and Humidity Range | 0~40℃ less than 80%RH (without condensation)  | 
| Storage Temperature and Humidity Range | -10~60℃ less than 70%RH (without condensation)  | 
| Power supply | LR03*3ea | 
| Dimensions | 216.9*86*Φ55mm | 
| Weight | 385g | 
| Optional accessory | MDU-100X/MDU-200X (display unit) |