DC Current Sensor
AC Current Sensor
Measuring Instruments

| SPECIFICATIONS | |
|---|---|
| Display | 3.5 digit LCD, max. reading of 1999 | 
| Measuring ranges | AC/DC 20A/200A | 
| Accuracy | 23℃±5℃ 80%RH or less (50/60Hz) Range/Resolution/Accuracy DC 20 A/0.01A/±1.5%rdg±3dgt DC200 A/0.1A/±2.0%rdg±3dgt(0~150.0A) DC200 A/0.1A/±3.0%rdg±3dgt(150.1~199.9A) AC 20 A/0.01A/±1.5%rdg±5dgt(50/60Hz) AC 20 A/0.01A/±2.0%rdg±5dgt(40~49Hz/61~500Hz) AC 200 A/0.1A/±2.0%rdg±5dgt(2.0~150.0A 50/60Hz) AC 200 A/0.1A/±2.5%rdg±5dgt(2.0~150.0A 40~49Hz/61~500Hz) AC 200 A/0.1A/±3.0%rdg±5dgt(150.1~199.9A 50/60Hz) AC 200 A/0.1A/±4.0%rdg±5dgt(150.1~199.9A 40~49Hz/61~500Hz)  | 
| Inside diameter of CT | φ30mm | 
| Withstanding voltage | AC 2000V/1 minute | 
| Over range indication | "OL" mark on LCD | 
| Data hold indication | "DH" mark on LCD | 
| Low battery indication | "B" mark on LCD | 
| Sampling rate | 2 times/sec | 
| Operating temperature | -10~50℃, < 80%RH (without condensation) | 
| Power supply | LR-44 battery×2 | 
| Dimension | 44(W)×148(H)×20(D)mm | 
| Weight | Approx. 100g | 