DC Current Sensor
AC Current Sensor
Measuring Instruments
CT Sensor | |
---|---|
CT | φ37mm |
Opening/closing of the jaw | AC2200V, 1 minute between the core of CT and CT outer case |
Withstanding voltage | AC2200V, 1 minute between the core of CT and CT outer case |
Measuring and Display Unit | |
---|---|
Measuring function | Leakage current, Harmonic current(Fundamental and third harmonics) |
Measuring method | CT clamp sensor |
Measuring range | 0-300μA/3mA/30mA(3range manual) |
Input frequency | 45-60Hz |
AC conversion | AC coupled true rms responding |
A/D conversion | Dual slope integration mode |
Display | LCD 3200 count max |
Sampling rate | 2 times/sec |
Over indication | "OL" mark on LCD |
Low battery Indication | "B" mark on LCD |
Data hold function | "DH" mark on LCD |
Auto power off | Approx.10 minutes later after power on |
Data output | DC 300mV full scale to each range(Waveform output) |
Power supply | 9V Alkaline battery 6LR61 × 1 |
Power consumption | Approx. 20mW |
Limitation of circuit voltage | Less than 500V AC |
Operating Temperature | 0~40℃, less than 80%RH, w/o condensation |
Storage Temperature | -10~60℃, less than 70%RH, w/o condensation |
Dimensions and Weight | Unit:/95(W)x160(H)x34(D)mm, 260gs CT:/135(W)x166(H)x61(D)mm, 1000gs |
AC Current | |
---|---|
Accuracy | (23℃±5℃, less than 80%RH) Range/Resolution/Accuracy(45~65Hz) */Max.Input Current 300μA/100nA(0.1μA)/1.2%±8digit/40A rms 3mA/1μA(0.001mA)/1.2%±8digit/40A rms 30mA/10μA(0.01mA)/1.2%±8digit/40A rms * Accuracy specified: More than 4% harmonics are necessary against fundamental harmonics |
Harmonic Current Measurement (Fundamental and third harmonics) | |
---|---|
Measuring method | PLL method |
Minimum fundamental input | More than 3% of full scale in each range |
Accuracy | (1%±5digit)±(Basic accuracy of ACA) -(Error by neighboring harmonics) * Accuracy specified:More than 4% harmonics are necessary against fundamental harmonics |
Accessories | Soft case:1 Instruction manual:1 |