Clamp type AC Current Sensor/for Low Leakage Current Measurements(ZCT Series)

ZCT-340L Low leakage current measurements with noise proof, AC 0.01mA~20A, φ40mm

FEATURES

Best sensor for low leakage current measurements of less than 1mA.
Suitable for low current measurements of terminal feeder.
High sensitivity and least affection from external magnetic field.
SPECIFICATIONS
Rated Primary Current AC20A
Applicable Current 0.01mA~20A
Max. Capable Current 30A
Secondary Windings 2000Turns
CT Inside Diameter φ35mm
Applicable Frequency 10Hz~500Hz
Output Example 49.5mV±2%(10A/10Ω)
Recommendable Load Resistor 10Ω
Output Part Lead Wire: Approx. 200mm
Rated Voltage Less than AC600V for low voltage circuit of coated wires
Withstanding Voltage AC2200V/1 minute (between output terminal and CT)
Insulation Resistance More than 100MΩ by 500V insulation tester (between output terminal and CT)
Operation Temperature 0~60℃, less than 80%RH w/o condensation
Dimension 68×179×30mm
Weight Approx. 120g