直流電流センサ
交流電流センサ
計測器
Insulation resistance | |
---|---|
Rated voltage and effective measuring range | MIS-2D:125V-20MΩ,250V-50MΩ,500V-100MΩ MIS-3D:250V-50MΩ,500V-100MΩ,1000V-2000MΩ MIS-4D:125V-20MΩ,250V-50MΩ,1000V-2000MΩ |
Minimum measurable resistance at rated voltage | MIS-2D:0.125MΩ,0.25MΩ,0.5MΩ MIS-3D:0.25MΩ,0.5MΩ,1MΩ MIS-4D:0.125MΩ,0.25MΩ,1MΩ |
Rated current | 1mA+20%-0% |
Open circuit voltage | Rated voltage+30%-0% |
Short circuit current | <2mA |
SPECIFICATIONS | |
---|---|
Withstanding Voltage | AC 3700V, 1 minute (Between input terminal and outer case) |
Storage Temperature | -10℃~60℃, 80% RH max (Without condensation) |
Operating Temperature | 0℃ to 40℃, 80% RH max. (Without condensation) |
Power supply | 1.5V (AM-3, R6 or "AA" size)×6 |
Weight | Approx. 330g (Excluding batteries) |
Safety standard | IEC 61010-1 Installation categoryⅡ, 600V phase to earth. |
Size | 170(W)×105(D)×54(H)mm |
Accessories | Line test lead:1 Earth test lead:1 Batteries:6 Test lead case:1 Belt:1 Instruction manual:1 |
Display | 3.5 digit LCD with bar graph display, max. reading of 3200 count and annunciators |
Optional accessory | Test probe with remote switch |
Data hold indication | "DH" mark on LCD readout |
Low battery indication | Battery mark on LCD readout |
E.M.C. standard | EN 61326 |
Function | Insulation resistance, AC voltage |
Constructional standard | In accordance with IEC1557-2 or JIS C1302 (1994) |
Overload protection | 120% of the highest nominal output voltage (10 sec.) |
Battery check | DC 6.3V~9.5V |
Temperature characteristics(0~40℃) | ±5% rdg of specified accuracy |
Accuracy(DC 125V) | Measuring range:0~20MΩ~OL(∞) First effective range:0.02MΩ~10MΩ<±5%rdg Second effective range:0.01MΩ~0.02MΩ 10MΩ~20MΩ <±10%rdg Other range:100MΩ~OL(∞) :Accuracy is not Specified |
Accuracy(DC 250V) | Measuring range:0~50MΩ~OL(∞) First effective range:0.05MΩ~20MΩ<±5%rdg Second effective range:0.02MΩ~0.05MΩ 20MΩ~100MΩ <±10%rdg Other range:100MΩ~OL(∞) :Accuracy is not Specified |
Accuracy(DC 500V) | Measuring range:0~100MΩ~OL(∞) First effective range:0.1MΩ~50MΩ <±5%rdg Second effective range:0.05MΩ~0.1MΩ 50MΩ~100MΩ <±10%rdg Other range:100MΩ~OL(∞) :Accuracy is not Specified |
Accuracy(DC 1000V) | Measuring range:0~2000MΩ~OL(∞) First effective range:2MΩ~1000MΩ <±5%rdg Second effective range:1MΩ~2MΩ 1000MΩ~2000MΩ <±10%rdg Other range:100MΩ~OL(∞) :Accuracy is not Specified |
AC voltage (50/60Hz) | Range/Accuracy/Input impedance/Maximum input voltage AC 600V/±2.5% of full scale/Approx. 20M/AC 600V rms |
Response time | Less than 5 sec. (Auto ranging) |
Infinity indication | "OL" mark on LCD readout (Over 3200 count) |