method of operation
SPECIFICATIONS 1 | ||||
---|---|---|---|---|
Rated Measuring Voltage | PVL 500V | PVH 1000V | ||
Effective Max. Display | 100MΩ | 2000MΩ | ||
Center Displayed Value | 2MΩ | 50MΩ | ||
First Effective Range | More than 0.1 MΩ, less than 50MΩ | More than 2 MΩ, less than 1000MΩ | ||
Tolerance | Less than ±5% | |||
Second Effective Range | More than 0.05MΩ, less than 0.1MΩ | More than 1MΩ, less than 2MΩ | ||
More than 50MΩ, less than 100MΩ | More than 1000MΩ, less than 2000MΩ | |||
Tolerance | Less than ±10% | |||
Judge Deterioration Point | Deterioration point will be displayed on LCD in case of insulation resistance less than 1MΩ. ※Only during measurement of PV panels, indicate P or N phase and or between modules. |
SPECIFICATIONS 2 | |
---|---|
Measuring Range (DC Voltage) | DC 0~999V |
Display | 3 1/2 digit LCD, max. reading of 3200 with bar graph and annunciator |
Display Ranges | 3.200/32.00/320.0/3200(4 range auto) |
Minimum Resolution | 0.001 MΩ |
Operating Temperature | 0℃~40℃, less than 80%RH (w/o condensing) |
Power Supply | 1.5V (AA size, LR6) alkali battery x 6 pcs. |
Dimension & Weight | 170(W)x105(D)x52(H)mm, approx. 350g (without batteries) |
Standard Accessories | line test lead ×2, earth test lead ×1, batteries ×6, Carrying case ×1, instruction manual ×1 |
OTHER FUNCTION | |
---|---|
Over Range Indication | ”OL” mark on LCD |
Low Battery Indication | ” B ” mark on LCD, less than 7.8±0.2V |
Data Hold Indication | ”DH” mark on LCD |
Auto Power Off | automatically power off, approx. 10 minutes after power on (To release this function, put the rotary switch to off and power on again) |
Back Light Function | By pressing back light switch (LIGHT) once, back light will appear and will become off by pressing once more. This back light will become off automatically 10 minutes after lighting on. |
Open Circuit Discharge | cancel the stored DC electrical charge by automatic discharging function, can confirm discharge by becoming HV lamp off. |