DC Current Sensor
AC Current Sensor
Measuring Instruments

| Insulation resistance | |
|---|---|
| Rated voltage and effective measuring range | MIS-1A:50V-10MΩ,125V-20MΩ,250V-50MΩ MIS-2A:125V-20MΩ,250V-50MΩ,500V-100MΩ MIS-3A:125V-20MΩ,250V-50MΩ,1000V-2000MΩ MIS-4A:250V-50MΩ,500V-100MΩ,1000V-2000MΩ  | 
| Center scale | MIS-1A:0.2MΩ/0.5MΩ/1MΩ MIS-2A:0.5MΩ/1MΩ/50MΩ MIS-3A:0.5MΩ/1MΩ/2MΩ MIS-4A:1MΩ/2MΩ/50MΩ  | 
| Minimum measurable resistance at rated voltage | MIS-1A:0.05MΩ,0.125MΩ,0.25MΩ MIS-2A:0.125MΩ,0.25MΩ,0.5MΩ MIS-3A:0.125MΩ,0.25MΩ,1MΩ MIS-4A:0.25MΩ,0.5MΩ,1MΩ  | 
| Rated current | 1mA+20%-0% | 
| Open circuit voltage | Rated voltage+30%-0% | 
| Short circuit current | <2mA | 
| SPECIFICATIONS | |
|---|---|
| Withstanding Voltage | AC 3700V, 1 minute (Between input terminal and outer case) | 
| Storage Temperature | -10℃~60℃, 80% RH max (Without condensation) | 
| Operating Temperature | 0℃ to 40℃, 80% RH max. (Without condensation) | 
| Power supply | 1.5V (AM-3, R6 or "AA" size)×6 | 
| Weight | Approx. 330g (Excluding batteries) | 
| Safety standard | IEC 61010-1 (2001) Installation categoryⅡ, 600V phase to earth.  | 
| Size | 170(W)×105(D)×54(H)mm | 
| Accessories | Line test lead:1 Earth test lead:1 Batteries:6 Test lead case:1 Belt:1 Instruction manual:1  | 
| Optional accessory | Test probe with remote switch | 
| E.M.C. standard | EN 61326 | 
| Function | Insulation resistance, AC voltage, Battery check | 
| Meter movement | 100μA, 870Ω, Taut band meter | 
| Constructional standard | In accordance with IEC1557-2 or JIS C1302 (1994) | 
| Overload protection | 120% of the highest nominal output voltage (10 sec.) | 
| Battery check | DC 6.3V~9.5V | 
| Low battery limit | DC 6.3V | 
| Temperature characteristics(0~40℃) | ±5% rdg of specified accuracy | 
| Accuracy(DC 50V) | First effective range:0.01MΩ~5MΩ ±5%rdg、10MΩ~50MΩ ±30%rdg Second effective range:  | 
| Accuracy(DC 125V) | First effective range:0.02MΩ~10MΩ ±5%rdg Second effective range:0.01MΩ~0.02MΩ 10MΩ~20MΩ ±10%rdg,10MΩ~100MΩ ±30%rdg  | 
| Accuracy(DC 250V) | First effective range:0.05MΩ~20MΩ ±5%rdg Second effective range:0.02MΩ~0.05MΩ 20MΩ~50MΩ ±10%rdg,50MΩ~100MΩ ±30%rdg  | 
| Accuracy(DC 500V) | First effective range:0.1MΩ~50MΩ ±5%rdg Second effective range:0.05MΩ~0.1MΩ 50MΩ~100MΩ ±10%rdg  | 
| Accuracy(DC 1000V) | First effective range:2MΩ~1000MΩ ±5%rdg Second effective range:1MΩ~2MΩ 1000MΩ~2000MΩ ±10%rdg  | 
| AC voltage (50/60Hz) | Range/Accuracy/Input impedance/Maximum input voltage AC 600V/±2.5% of full scale/Approx. 1.5M/AC 600V rms  |