DIGITAL INSULATION RESISTANCE TESTER
PRODUCTS
Model MIS-2D
Model MIS-3D
Model MIS-4D
125V / 20M Ω
250V / 50M Ω
500V / 100M Ω
250V / 50M Ω
500V / 100M Ω
1000V / 2000M Ω
125V / 50M Ω
500V / 100M Ω
1000V / 2000M Ω
(CE)
FEATURES
The big digital and bar graph LCD display with back light enabled easy observation. Especially useful when working in dark place.
Hand free and continuous measurements with custom made switch.
Safe design with built in automatic discharging function for any capacitors present in the circuit.
The voltage in the circuit or capacitor can be checked by warning lamp for safe insulation measurements.
Data hold and auto power off function.
Compact, light weight and heavy duty rugged case.
≫
Show detail
≫
Catalog PDF
DETAIL
SPEC
Measuring Ranges and Technical Data
Insulation resistance
Model
MIS-2D
MIS-3D
MIS-4D
Rated voltage and
effective measuring range
125V-20MΩ
250V-50MΩ
500V-100MΩ
250V-50MΩ
500V-100MΩ
1000V-2000MΩ
125V-20MΩ
250V-50MΩ
1000V-2000MΩ
Minimum measurable
resistance at
rated voltage
0.125MΩ
0.25MΩ
0.5MΩ
0.25MΩ
0.5MΩ
1MΩ
0.125MΩ
0.25MΩ
1MΩ
Rated current
1mA+20%-0%
Open circuit voltage
Rated voltage+30%-0%
Short circuit current
<2mA
SPECIFICATIONS
Function
:
Insulation resistance, AC voltage
Display
:
3.5 digit LCD with bar graph display, max. reading of 3200 count and annunciators
Safety standard
:
IEC 61010-1
Installation categoryⅡ, 600V phase to earth.
E.M.C. standard
:
EN 61326
Response time
:
Less than 5 sec. (Auto ranging)
Data hold indication
:
"DH" mark on LCD readout
Infinity indication
:
"OL" mark on LCD readout (Over 3200 count)
Constructional standard
:
In accordance with IEC1557-2 or JIS C1302 (1994)
Withstanding voltage
:
AC 3700V, 1 minute (Between input terminal and outer case)
Overload protection
:
120% of the highest nominal output voltage (10 sec.)
Battery check
:
DC 6.3V~9.5V
Low battery indication
:
Battery mark on LCD readout
Operating temperature
:
0℃ to 40℃, 80% RH max. (Without condensation)
Temperature characteristics
(0~40℃)
:
±5% rdg of specified accuracy
Storage temperature
:
-10℃~60℃, 80% RH max (Without condensation)
Power supply
:
1.5V (AM-3, R6 or "AA" size)×6
Size
:
170(W)×105(D)×54(H)mm
Weight
:
Approx. 330g (Excluding batteries)
Accessories
:
Line test lead
:
1
Earth test lead
:
1
Batteries
:
6
Test lead case
:
1
Belt
:
1
Instruction manual
:
1
Optional accessory
:
Test probe with remote switch
Accuracy
Rated voltage
DC 125V
DC 250V
DC 500V
DC 1000V
Measuring range
0~20MΩ~OL(∞)
0~50MΩ~OL(∞)
0~100MΩ~OL(∞)
0~2000MΩ~OL(∞)
First effective range
0.02MΩ~10MΩ
<±5%rdg
0.05MΩ~20MΩ
<±5%rdg
0.1MΩ~50MΩ
<±5%rdg
2MΩ~1000MΩ
<±5%rdg
Second effective range
0.01MΩ~0.02MΩ
10MΩ~20MΩ
<±10%rdg
0.02MΩ~0.05MΩ
20MΩ~100MΩ
<±10%rdg
0.05MΩ~0.1MΩ
50MΩ~100MΩ
<±10%rdg
1MΩ~2MΩ
1000MΩ~2000MΩ
<±10%rdg
Other range
100MΩ~OL(∞) :Accuracy is not Specified
AC voltage (50/60Hz)
Range
Accuracy
Input impedance
Maximum input voltage
AC 600V
±2.5% of full scale
Approx. 20M
AC 600V rms