Digital Harmonics Clamp Tester
PRODUCTS
|
AC Leakage Current, 400mA/4A/40A/300A Harmonics current, 40mmφCT
Model HWT-300
|
FEATURES
- Harmonics and leakage current measurements.
- Data-hold and Auto power off function.
- The best tool to measure up to 25th harmonics current in the field.
- Analog data output function.
|
|
≫ Show detail
≫ Catalog PDF
|
DETAIL
SPEC
| SPECIFICATIONS |
| Measuring method |
: |
Dual slope integration mode |
| Input frequency |
: |
45Hz~65Hz |
| Display |
: |
3.5 digit LCD, max. reading of 3999 |
| Jaw opening capability |
: |
40mmφ |
| Data output |
: |
DC 100mV for full scale(400mA/4A/40A ranges)
DC75mV for full scale(300A range)
Accuracy ±1% of full scale
Output impedance 10kΩ or less
|
| Over range indication |
: |
"OL" mark on LCD |
| Low battery indication |
: |
"Battery" mark on LCD |
| Data hold indication |
: |
"DH" mark on LCD |
| Sampling |
: |
2 times/sec |
| Withstanding voltage |
: |
AC 2200V 1 minute max. (Between the core of CT and outer case) |
| Affection of magnetic field |
: |
3mA or less (at 100A nearby conductor) |
| Limitation of circuit voltage |
: |
Less than AC 600V |
| Operating temperature |
: |
0℃ to 40℃, 80%RH max. ( Without condensation ) |
| Storage temperature |
: |
-10℃ to 60℃, 70%RH max. ( Without condensation ) |
| Power supply |
: |
1.5V (AM-4, LR03 or AAA)×3 or AC adaptor(Option) |
| Power consumption |
: |
13mA |
| Auto power off |
: |
Approx.10 minutes later after power on |
| Battery life |
: |
Approx.50 hours continuous |
| Size |
: |
68(W)×207(H)×33(D)mm |
| Weight |
: |
Approx. 430g |
| Accessories |
: |
| Hard carrying case |
: |
1 |
| Instruction manual |
: |
1 |
| Batteries |
: |
3 |
|
| Option |
: |
AC adaptor |
| Accuracy |
: |
(23℃±5℃, 80%RH or less)
| Range |
Resolution |
Accuracy |
Max.input |
| AC 400mA |
0.1mA |
±1.0% rdg ± 8 dgt |
AC 300Arms |
| AC 4A |
1mA |
| AC 40A |
10mA |
| AC 300A |
100mA |
±1% of full scale |
|
|
| Harmonics mode |
| Measuring method |
: |
Synchronous filter |
| Measurable harmonics |
: |
Fundamental to 25th harmonics |
| Minimum input |
: |
More than 5% of full scale in each range |
| Accuracy |
: |
±1%rdg±5 dgt (23℃±5℃) |
Error by neighboring harmonics
| Harmonics |
Neighboring harmonics component ratio against current or voltage |
Typical accuracy |
| 5th |
65% |
±3%rdg ± 5dgt |
| 7th |
41% |
±3.5%rdg ± 5dgt |
| 11th, 13th |
20% |
±4%rdg ± 5dgt |
| 15th~25th |
10% |
±5%rdg ± 5dgt |
|
|